Housing |
19″-version for mounting in control cabinets 3 U / 42 HP
dimensions:W 213.0 mm (8.39 in.) H 128.4 mm (5.06 in.) D 230.0 mm (9.05 in.)
material:stainless steel / polyester / silicone / glass / diverse plastics
protection:front IP40 / rear IP20 (mains supply secured against accidental touching)
|
Display |
LCD graphic display black on white (240 x 128 pixel), LED background illuminated
|
Operation |
18-button keyboard
|
System clock |
accuracy approx. 1 minute/month (battery life approx. 15 years)
|
LED |
1 LED (green):power on
1 LED (red-flashing):system failure
3 LEDs (yellow):alarm I, II, III
|
Data logger |
4 parallel measuring values
(ring buffer with approx. 25,000 data points x 4)
(interval: 1/second – 1/hour)
|
Sensor-inputs |
1 – 4 for optek photometric sensors
|
Sensor-inputs
(explosion proof) |
optional:1 – 4 for optek photometric sensors (intrinsic safe)
|
mA-inputs |
optional:
2 x 4 – 20 mA (functionally galvanically isolated)
accuracy: < 0.5 %
resolution: < 0.05 %
load: < 200 Ohm
|
Remote-inputs |
optional:7 x 24 V (19 … 29 V DC),
typically 6.0 mA for remote range setting, remote zero, remote hold
|
Profibus® PA interface |
optional:
Profibus® PA profile, version 3.01, amendment 2
|
FOUNDATION™ Fieldbus interface |
optional:
FOUNDATION™ Fieldbus H1 (IEC 61158-2)
|
Sensor lamp-outputs |
1 or 2 lamp supply for optek photometric sensors
4.5 … 8.5 V DC
|
mA-outputs |
2 or 4 x 0/4 – 20 mA (NAMUR)
(functionally galvanically isolated)
accuracy: < 0.5 %
resolution: < 0.05 %
load: < 600 Ohm
|
Relay-outputs |
3 independent software-configurable relay contacts
0 – 50 V AC, 0 – 75 V DC, 0 – 2 A
for alarm or status feedback
initiation delay configurable: 0 – 999 sec.
|
Failsafe-output |
1 SPDT contact to alarm in case of lamp or system failure (active)
0 – 50 V AC, 0 – 75 V DC, 0 – 2 A
|
Serial communication |
RS232 bi-directional interface on front panel (with software package optek PC-transfer)
upload / download of configuration, download of data logger content
|
Cable lengths
(sensor) |
2, 3, 5, 10, 15, 20, 30 … 100 m
(7, 10, 16, 33, 49, 66, 98 … 328 ft)
cable length > 100 m on request up to 1,000 m (3,280 ft)
sensors: AS56 / AS16: max: 50 m
sensors: ASD: 2, 3, 5 or 10 m (7, 10, 16 or 33 ft.)
|
Power supply |
115 / 230 V AC, selectable (93.5 – 132 / 187 – 264 V AC, 47 – 64 Hz) or 24 V AC / DC (AC: 20.4 – 26.4 V AC, 47 – 64 Hz; DC: 20.4 – 28.8 V DC)
power consumption:< 50 VA
|
Ambient conditions |
temperature during operation (no direct sunlight):
converter: -10 – 55 °C (14 – 131 °F)
with optional stainless steel housing S19-42 (IP65): -20 – 45 °C ( -4 – 113 °F)
with optional plastic housing B19-42 (IP66): -10 – 40 °C (14 – 104 °F) (C4000 only)
with optional Ex D housing (IP65): -20 – 40 °C ( -4 – 104 °F) (C4000 only)
temperature during transport (no direct sunlight): -20 – 70 °C ( -4 – 158 °F)
|
Software languages |
English, German, French, Spanish, Dutch, Portuguese, Russian
|