VisiFerm mA Optik Çözünmüş Oksijen Sensörü

Precise DO Measurement,
Less Frequent Calibration

Hamilton extensively investigated the effects of the bioprocess on measurement accuracy. Analysis of sensor aging, chemical attack, and physical damage yielded the necessary understanding for development of a next generation VisiFerm which reduces the need for frequent calibration.

Learn the Sources of DO Measurement Error

Get Hamilton’s whitepaper on potential sources of optical dissolved oxygen measurement error and ways to mitigate the impact.

 

Açıklama

Hamilton extensively investigated the effects of the bioprocess
on measurement accuracy. Analysis of sensor aging, chemical
attack, and physical damage yielded the necessary understanding
for development of a next generation VisiFerm which reduces
the need for frequent calibration.

Part # / Ref Sensor Family a-length Sensor Cap
10070760-1111
VisiFerm mA
120 mm
ODO Cap H3
10070760-1121
VisiFerm mA
120 mm
ODO Cap H4
10070760-1211
VisiFerm mA
160 mm
ODO Cap H3
10070760-1221
VisiFerm mA
160 mm
ODO Cap H4
10070760-1311
VisiFerm mA
215 mm
ODO Cap H3
10070760-1321
VisiFerm mA
215 mm
ODO Cap H4
10070760-1411
VisiFerm mA
325 mm
ODO Cap H3
10070760-1421
VisiFerm mA
325 mm
ODO Cap H4
10070760-1511
VisiFerm mA
425 mm
ODO Cap H3
10070760-1521
VisiFerm mA
425 mm
ODO Cap H4

Teknik Özellikler

Sensor Family VisiFerm mA
a-length 120 mm
Electrical Connector M12
Measurement Principle Oxygen dependent luminescence quenching, temperature compensation
Measuring Range 4 ppb to 25 ppm (DO) 0 to 62.85 %-vol or 0 to 300 %-sat
Accuracy at 25 °C 1 ± 0.05 %-vol: 21 ± 0.2 %-vol; 50 ± 0.5 %-vol
Drift at Room Temperature < 1 % per week

Additional specifications

Sensor Cap ODO Cap H3
Electrolyte None
Oxygen Consumption None
Temperature Sensor NTC 22 kOhm
Configurable Values DO: mbar; %-sat; %-vol; µg/l; mg/l; ppb/ppm (gas); ppb/ppm (dissolved oxygen); Temperature: °C
Diameter 12 mm
Process Connection PG13,5
Wetted Parts Stainless Steel 1.4435
EPDM (Ethylene propylene elastomer)
VMQ (Silicone elastomer)
See compliance details in Material Specification document
Surface Quality of Steel Ra < 0.4 µm (N5)
4 to 20 mA Current Range 3.5 to 22 mA
Analog Interface 1 Two wire sink 4 to 20 mA for DO, programmable
4 to 20 mA Accuracy < 0.3 % current value + 0.05 mA
Transmitting Power Max. 4 dBm
Digital Interface Bluetooth 5
HART Version 7.0
Operating Voltage 18 to 30 VDC
Serial Number Yes
Response Time t98% < 30 s at 25 °C
Certificate Yes, with measured current in air
ATEX Approval ll 1 G Ex ia llC T3/T4/T5/T6 Ga; II 1 D Ex ia IIIC T135°C Da
IECEx Approval ll 1 G Ex ia llC T3/T4/T5/T6 Ga; II 1 D Ex ia IIIC T135°C Da
Autoclavable Yes
CIP Yes
Steam Sterilizable Yes, max. Temperature 140 °C
Operating Temperature Range -20 to 140 °C; the sensor provides no DO reading above 85 °C
Pressure Range bar g -1 to 12 bar
Protection Rating IP 68
Required Flow None

Teknik Döküman

10082866-02_VisiFerm-Optical-DO-Brochure-LR